abstract:
Original paper

Radiation Dose Associated with Multi-Detector 64-Slice Computed Tomography Brain Examinations in Khartoum State, Sudan

Ali Elmahdi
,
Mohamed M. Abuzaid
,
Esameldeen Babikir
,
Abdelmoneim Sulieman

Pol J Radiol, 2017; 82: 603-606
Online publish date: 2018/01/18
View full text Get citation
 
PlumX metrics:
Background: Radiation exposure due to computed tomography (CT) has become an important issue, as the number of CT examinations has been increasing worldwide. Radiation doses associated with CT are higher in comparison to other imaging procedures. CT-related radiation doses should be monitored and controlled in order to ensure reduction of radiation exposure and optimization of image quality. The aim of this study was to evaluate radiation doses in adult patient who underwent routine CT brain examinations, and to assess how CT scanning protocols affect patient doses in practice.

Material/Methods: A total of 118 patients underwent brain CT at two radiology departments equipped with 64-slice CT scanners, Khartoum, Sudan. Patient doses regarding weighted CT dose index (CTDIw) and dose length product (DLP) values were recorded. Quality control tests were performed for both scanners.

Results: The mean CTDIw values ranged from 62.9 to 65.8 mGy, DLP values ranged from 1003.7 to 1192.5 mGy, and the effective dose varied from 2.4 to 3.7 mSv.

Conclusions: Patient doses in this study was higher compared to previous research, suggesting that patients exposed to unnecessary radiation. Therefore, optimization of radiation doses with the use of specified imaging protocols, well-documented indications for CT, training of technicians, and quality control programs will reduce the necessary radiation doses. Establishment of the diagnostic reference level is recommended for further dose reduction.
keywords:

Multidetector Computed Tomography • Radiation Dosage • Tomography, X-Ray Computed