The assessment of diagnostic medical images using 10-bit grayscale – fact or myth?

Robert Chrzan, Andrzej Urbanik

© Pol J Radiol 2018; 83: e127-e132
Online publish date: 2018/04/05
View full text
Get citation
ENW
EndNote
BIB
JabRef, Mendeley
RIS
Papers, Reference Manager, RefWorks, Zotero
AMA
APA
Chicago
Harvard
MLA
Vancouver
 
Purpose
The aim of the study was to find tests for verification of 10-bit grayscale support for radiological workstation hardware/operating system and for DICOM viewer. Also, we wanted to perform the tests on available workstations.

Material and methods
The following tests were selected to verify 10-bit grayscale support: workstation hardware and operating system tests using DirectX SDK 10BitScanout10.exe and NEC 10bitdemo.exe programs, DICOM viewer tests using TG18-MP and TG18-QC patterns. The tests were performed on 14 radiological workstations in a local department of radiology and on 4 radiological workstations during the technical exhibition at the ECR (European Congress of Radiology) in 2017.

Results
Only 3 out of 14 local workstations passed the hardware and operating system 10-bit grayscale support test. Only one workstation passed the DICOM viewer 10-bit grayscale support test.

Conclusions
Hardware/operating system and DICOM viewer 10-bit grayscale support should be tested on every radio­logical workstation even if compatibility is stated by the manufacturer. Existing medical diagnostic display quality requirements should be adjusted so that 10-bit support for DICOM viewers is only a desirable option.

keywords:

data display, diagnostic imaging, quality control

references:
Minister Zdrowia. Obwieszczenie Ministra Zdrowia z dnia 26 kwietnia 2013 r. w sprawie ogłoszenia jednolitego tekstu rozporządzenia Ministra Zdrowia w sprawie warunków bezpiecznego stosowania promieniowania jonizującego dla wszystkich rodzajów ekspozycji medycznej. Dz.U. z 2013 poz. 1015. Available from: http://dziennikustaw.gov.pl/DU/2013/1015/1 [in Polish].
Minister Zdrowia. Rozporządzenie zmieniające rozporządzenie w sprawie warunków bezpiecznego stosowania promieniowania jonizującego dla wszystkich rodzajów ekspozycji medycznej. Dz.U. z 2015 poz. 2040. Available from:
http://dziennikustaw.gov.pl/DU/2015/2040/1 [in Polish].
Minister Zdrowia. Obwieszczenie Ministra Zdrowia z dnia 3 kwietnia 2017 r. w sprawie ogłoszenia jednolitego tekstu rozporządzenia Ministra Zdrowia w sprawie warunków bezpiecznego stosowania promieniowania jonizującego dla wszystkich rodzajów ekspozycji medycznej. Dz.U. z 2017 poz. 884. Available from: http://dziennikustaw.gov.pl/DU/2017/884/1 [in Polish].
Sekcja Inżynierii Klinicznej PLTR. Wymagania dotyczące stanowiska do interpretacji (stanowiska opisowego) dla radiologii cyfrowej. 2010. Available from: http://los.net.pl/download.php?id=55 [in Polish].
EUREF. European protocol for the quality control of the physical and technical aspects of mammography screening. In: European guidelines for quality assurance in breast cancer screening and diagnosis Fourth edition. 2006. Available from: http://www.euref.org/downloads?download=26:physico-technical-protocol
DIN Deutsches Institut für Normung. DIN 6868-157 Image quality assurance in diagnostic X-ray departments – Part 157: X-ray ordinance acceptance and constancy test of image display systems in their environment. Available from: http://www.din.de/en/getting-involved/standards-committees/nar/standards/wdc-beuth:din21:216064441
The Institute of Physics and Engineering in Medicine. Report 91 Recommended Standards For The Routine Performance Testing of Diagnostic X-Ray Imaging Systems. Available from: https://www.ipem.ac.uk/ScientificJournalsPublications/IPEMReportSeries/AvailablePublications.aspx
American Association of Physicists in Medicine (AAPM) Task Group 18. Assessment of display performance for medical imaging systems. 2005. Available from: http://www.aapm.org/pubs/reports/or_03.pdf
Samei E, Badano A, Chakraborty D, Compton K et al.: Assessment of display performance for medical imaging systems: Executive summary of AAPM TG18 report. Med. Phys, 2005; 32(4): 1205–25
Japan Industries Association of Radiological Systems (JIRA). JESRA-X-0093 Quality Assurance (QA) Guideline for Medical Imaging Display Systems. Available from: http://www.jira-net.or.jp/e/information/file2/JESRA_X-0093_A-2010_e.pdf
nVidia. 30-Bit Color Technology for NVIDIA® Quadro®. 2009. Available from: https://www.nvidia.com/docs/IO/40049/TB-04701-001_v02_new.pdf
nVidia. 10 and 12-bit grayscale technology. 2013. Available from: ttps://www.nvidia.com/content/quadro_fx_product_literature/TB-04631-001_v05.pdf
Advanced Micro Devices. AMD’s 10-bit Video Output Technology. 2008. Available from: https://www.amd.com/Documents/10-Bit.pdf
Tabár LK, Nińo AS, Schreiber E et al: Three digital mammography display configurations: Observer performance in a pilot ROC Study. In: Martí J, Oliver A, Freixenet J, Martí R (eds.). Digital mammography. IWDM 2010. Lecture Notes in Computer Science, vol. 6136. Springer, Berlin, Heidelberg.
Kimpe T, Tuytschaever T. Increasing the number of gray shades in medical display systems – how much is enough? J Digit Imaging 2007; 20: 422-432.
 
Quick links
© 2018 Termedia Sp. z o.o. All rights reserved.
Developed by Bentus.
PayU - płatności internetowe