Transcatheter retrieval of an Amplatzer Vascular Plug

Yozo Sato
,
Yoshitaka Inaba
,
Takaaki Hasegawa
,
Shinichi Murata
,
Yuki Kimbara
,
Yoshiki Senda

© Pol J Radiol 2018; 83: e133-e136
Online publish date: 2018/04/08
Article file
Get citation
 
PlumX metrics:
 

Introduction

An Amplatzer Vascular Plug (AVP; Abbott Vascular, Santa Clara, California) is a useful embolic device that can be precisely deployed in medium to large vessels with high resistance to migration [1]. The AVP has developed from a single device to a group of four models (AVP, AVP II, AVP III, and AVP 4) that can be used for various types of anatomy, haemodynamics, and lesion entities. The original AVP was designed as a permanent occluding device derived from the Amplatzer Septal Occluder and Amplatzer Duct Occluder [1]. These Amplatzer devices have a self-expanding feature with sufficient radial force on the vessel wall to minimise movement; however, migration of the Amplatzer occluder in cardiac interventions has been reported as a relatively rare but major complication [2-4].

Case report

This case report of transcatheter retrieval of AVP migration to the proper hepatic artery was exempt from institutional review board approval. A 59-year-old woman was referred for the treatment of advanced pancreatic body cancer with paraaortic lymph node metastases. Enhanced computed tomography revealed shrinkage of primary tumour and lymph node metastases after systemic chemotherapy. No abnormal uptake of lesions in [18F]fluorodeoxyglucose-positron emission tomography/computed tomography (FDG-PET/CT) was observed. After seven cycles of systemic chemotherapy, distal pancreatectomy with en bloc celiac axis resection (DP-CAR) was planned as curative treatment. Therefore, preoperative embolisation of the common hepatic artery (CHA) for arterial redistribution was employed to prevent postoperative ischaemic complications.
Angiography using a 4-F diagnostic catheter revealed the left gastric artery (LGA) arising from the aorta and the left hepatic artery arising from the LGA. Celiac arteriography revealed CHA with a maximum diameter of 4.2 mm (Fig. 1A); a 4-F catheter was advanced to the gastroduodenal artery. A 5-F guiding sheath (Flexor® Ansel; Cook Medical, Bloomington, Indiana) was advanced to the CHA along with a 4-F catheter, and a 6-mm AVP II was deployed at the mid-portion of the CHA (Fig. 1B). Although the AVP was completely expanded, it did not move, even if force was exerted on the delivery cable. However, the AVP migrated to the proximal portion of the proper hepatic artery approximately 30 seconds after detachment (Fig. 2A). Although migrated AVP retrieval using a goose neck snare (Amplatz GooseNeck Snare; Medtronic, Minneapolis, Minnesota) was attempted, it was impossible to retrieve it into the 5-F guiding sheath because the AVP delivery screw became stuck within the sheath edge (Fig. 2B-C). However, the guiding sheath with snaring the AVP was gradually pulled to the proximal side and the guiding sheath could be directed to the splenic artery. Therefore, the AVP was delivered to the splenic artery, which was planned to be resected in DP-CAR. Finally, a 10-mm AVP II was redeployed at the proximal portion of the CHA, and complete occlusion was achieved (Fig. 3A-B). The distal portion of the splenic artery was visualised via the collateral blood flow from the LGA. Although mild-spleen infarction occurred, DP-CAR was successfully performed three days after embolisation.

Discussion

The utility of CHA preoperative embolisation for DP-CAR has been reported [5, 6]; AVP, which has a low risk of migration, can be an optimal embolic device. The incidence of device migration (including an Amplatzer Septal Occluder) associated with transcatheter occlusion of atrial septal defects was reported to be from 0.55% to 3.5% [2, 3]. The commonest cause of migration was the use of an undersized device. AVP II was completely expanded due to an undersized device after deployment in the case presented here; hence, the AVP should have been retrieved before detachment. Although transcatheter retrieval using a goose neck snare and/or a basket catheter was recommended, the technical success rate was reported to be from 50% to 75%, and surgical retrieval was required in failure cases [2-4]. In the present case, the migrated AVP could be snared; however, it was impossible to retrieve it into the 5-F guiding sheath. Because there was no larger diameter sheath at that time, there was no other option but to deliver the AVP to the splenic artery, which was planned to be resected in DP-CAR. It may be essential to prepare larger-sized sheaths that enable AVP retrieval without it becoming stuck in the delivery screw. However, when AVP retrieval is not possible, delivery to the other arteries having lesser influence might be an alternate technique.

Conclusions

This report presented a case of transcatheter retrieval of AVP migration to the proximal portion of the proper hepatic artery. Although migrated AVP retrieval using a goose neck snare was attempted, it was impossible to retrieve it into the 5-F guiding sheath. Therefore, the AVP was delivered to the splenic artery, which was planned to be resected in DP-CAR. When AVP retrieval is not possible, delivery to the other arteries having lesser influence might be an alternate technique.

Conflict of interest

The authors report no conflict of interest.

References

1. Wang W, Li H, Tam MD, et al. The amplatzer vascular plug: a review of the device and its clinical applications. Cardiovasc Intervent Radiol 2012; 35: 725-740.
2. Moore J, Hegde S, El-Said H, et al. Transcatheter device closure of atrial septal defects: a safety review. JACC Cardiovasc Interv 2013; 6: 433-442.
3. Chessa M, Carminati M, Butera G, et al. Early and late complications associated with transcatheter occlusion of secundum atrial septal defect. J Am Coll Cardiol 2002; 39: 1061-1065.
4. Goel PK, Kapoor A, Batra A, et al. Transcatheter retrieval of embolized AMPLATZER Septal Occluder. Tex Heart Inst J 2012; 39: 653-656.
5. Takasaka I, Kawai N, Sato M, et al. Preoperative microcoil embolization of the common hepatic artery for pancreatic body cancer. World J Gastroenterol 2012; 18: 1940-1945.
6. Ishikawa M, Kajiwara K, Fukumoto W, et al. Blood Flow Redistribution Using the AMPLATZER Vascular Plug 4 before Distal Pancreatectomy with en Bloc Celiac Axis Resection. J Vasc Interv Radiol 2016; 27: 285-286.
Copyright: © Polish Medical Society of Radiology This is an Open Access article distributed under the terms of the Creative Commons Attribution-Noncommercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0). License allowing third parties to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially.