ISSN: 1899-0967
Polish Journal of Radiology
Established by prof. Zygmunt Grudziński in 1926 Sun
Current issue Archive Manuscripts accepted About the journal Editorial board Abstracting and indexing Contact Instructions for authors Ethical standards and procedures
Editorial System
Submit your Manuscript
SCImago Journal & Country Rank
1/2019
vol. 84
 
Share:
Share:
Gastrointestinal and abdominal radiology
Original paper

Perianal fistula imaging: a comparison between two-channel superficial Flex coil and eight-channel body coil

Nazlı Gülsüm Akyel
1
,
Kayıhan Akın
2
,
Dilek Kösehan
3
,
Aslı Köktener
3

1.
Cerrahpaşa Medical Faculty, Istanbul University, Istanbul, Turkey
2.
Lokman Hekim Akay Hospital, Ankara, Turkey
3.
Memorial Ankara Hospital, Ankara, Turkey
© Pol J Radiol 2019; 84: e430-e435
Online publish date: 2019/11/04
Article file
Get citation
 
PlumX metrics:
 

Introduction

Perianal fistula is an important health problem with devastating morbidity, including severe local pain, chronic perianal discharge, impairment of social activities, and incontinence [1]. It is defined as an epithelial-lined abnormal tract between the anal canal and the skin [2]. Its exact prevalence is not known, but it is estimated to be around 1 : 10 000. The male-to-female ratio is 2 : 1 [3].

Treatment is surgical, but due to missed infection sites, it has a tendency to recur; in some series it has had a recurrence rate of up to 25% [4]. Accurate preoperative assessment of primary perianal fistula and its relationship with sphincters, secondary tracks, and abscess formation is crucial for the establishment of a treatment plan. Besides its diagnostic purposes, magnetic resonance imaging (MRI) has become the gold standard method for perianal fistula imaging because it gives the most detailed information about the anatomical features of the fistula tract and its relationship with the anal sphincter muscles. By providing these data, MRI increases the rate of surgical success and decreases the recurrence rate of perianal fistula [2].

There are many publications about different magnetic resonance sequences with or without contrast material [5-9] and diffusion-weighted MRI [10-13]. Additionally, endoanal coils and body coils have previously been compared with each other [14] or by surgical correlation. In the literature, it is noteworthy that some of the studies used torso [15-17] or spine coils [18]. However, there has been no published literature comparing body coils and superficial coils so far. Therefore, in our present study, we aimed to investigate whether using a two-channel Flex-M superficial coil would give more information compared to the images obtained with an eight-channel phased-array body coil in perianal fistula imaging.

Material and methods

Patients

All patients with anal fistula were planned to be involved in our present study. The presence of malignancy was accepted as an exclusion criterion.

All fistulas were idiopathic except in one patient with Crohn’s disease. All patients were informed about the study, and the patients who agreed to be involved in the study were recruited accordingly. Our study was approved by the Local Ethics Committee, and informed consent was obtained from each study participant.

Imaging technique

Patients underwent 1.5 T MRI imaging (Achieva; Philips Medical Systems; Best, Netherlands) in the supine position. Each examination included sagittal fat-suppressed (FS) T2-weighted (T2W) (time of repetition/time of echo [TR/TE], 3500/80 ms), coronal FS T2W (TR/TE, 3500/80 ms), axial T1W (TR/TE, 550/7 ms), T2W (TR/TE, 4500/80 ms), FS T2W (TR/TE, 4500/80 ms), FS T1W (TR/TE, 550/7 ms), axial dynamic THRIVE (TR/TE, 4.6/2.2 ms), postcontrast axial, sagittal, and coronal FS T1W (TR/TE, 475-650/7 ms) images. Other imaging parameters included a slice thickness of 3.5 mm with an interslice gap of 1 mm, a matrix size of 175 × 385, and a field of view (FOV) of 240 × 240 mm.

Distal rectum and skin were included. Axial and coronal imaging was performed as oblique axial and oblique coronal parallel to the anal canal (true axial and true coronal). Postcontrast series were obtained following bolus intravenous injection of 0.1 mmol/kg MRI contrast agent (Gadoversetamid, Optimark, Covidien).

All MRI investigations were performed with a phased-array body coil. Additionally, an axial FS T2W sequence was gathered with Sense Flex-M coil (Achieva; Philips Medical Systems; Best, Netherlands) as a superficial coil. Patients were evaluated at a workstation. Afterwards, the axial FS T2W sequences that were previously obtained with Flex-M and body coils were evaluated separately by two experienced radiologists, who were blinded to the patients. Interpretations of the images by the two radiologists were made at different times. The radiologists were unaware of each other’s interpretations.

Flex and body coils were connected to the patient at the same time. The Flex coils were placed at the front and back of the patient’s anal region, and a body coil was wrapped around it.

The site of internal and external openings and the presence of any abscess and secondary tracts were evaluated. Cases were categorised according to Park’s [19] and St. James University [20] classification systems.

Statistical analysis

All statistical analyses were performed using the SPSS 21.0 Statistical Package Program for Windows (SPSS Inc., Chicago, IL, USA). Categorical variables were compared with the χ2 test. Intra- and inter-observer concordance were tested by Cohen’s κ test. P values < 0.05 were accepted as statistically significant.

Results

A total of 29 patients with 31 fistulas were included in our present study. Of the patients, 19 were male and 10 were female. The mean age of the study group was 42.8 years (age range: 19-68).

Reader 1. Twenty-seven external openings of the fistulas could be seen by MRI with body coil, whereas 29 external openings could be seen by MRI with Flex-M coil. Cohen’s κ value for MRI with body coil and Flex-M coil was 0.635 (p < 0.001).

By MRI with body coil, 19 internal openings of the fistulas could be seen, whereas 30 internal openings could be seen by MRI with Flex-M coil. Cohen’s κ value was 0.100 (p = 0.201) for the two techniques (Table 1). Six secondary fistula tracts were reported with body coil, whereas eight secondary fistula tracts were reported with Flex-M coil.

Table 1

Power of magnetic resonance imaging with Flex-M and body coil for visualisation of external opening, internal opening, secondary tract, and abscess related with perianal fistula

Reader 1Reader 2
Flex-M coilBody coilFlex-M coilBody coil
External opening
 Patients, n (%)29 (93.5)27 (87.1)27 (87.1)25 (80.6)
 Cohen’s κ value*0.635 (p < 0.001)0.763 (p < 0.001)
Internal opening
 Patients, n (%)30 (96.8)19 (61.3)28 (90.3)18 (58.1)
 Cohen’s κ value*0.100 (p = 0.201)0.110 (p = 0.361)
Secondary tract
 Patients, n (%)8 (25.8)6 (19.4)8 (25.8)6 (19.4)
 Cohen’s κ value*0.817 (p < 0.001)0.817 (p < 0.001)
Abscess
 Patients, n (%)6 (19.4)6 (19.4)6 (19.4)6 (19.4)
 Cohen’s κ value*1.000 (p < 0.01)1.000 (p < 0.01)

* Kappa values show concordance between MRI with Flex-M coil and MRI with body coil

Reader 2. Twenty-five external openings could be seen by MRI with body coil, whereas 27 external openings could be seen by MRI with Flex-M coil. Cohen’s κ value for MRI with body coil and Flex-M coil was 0.763 (p < 0.001).

Eighteen internal openings could be seen by MRI with body coil, whereas 28 internal openings could be seen by MRI with Flex-M coil. Cohen’s κ value for MRI with body coil and Flex-M coil was 0.110 (p = 0.361).

Six secondary fistula tracts were reported with body coil, whereas eight secondary fistula tracts were reported with Flex-M coil. Cohen’s κ value for MRI with body coil and Flex-M coil was 0.817 (p < 0.001).

There were six abscesses in six different patients, and all these abscesses were reported by both readers with both imaging techniques.

A second fistula was seen in two different patients. Neither reader could see the second fistulas by MRI with body coil. However, Reader 1 reported the second fistula of both patients and Reader 2 reported one of them by MRI with Flex-M coil.

Inter-observer variability for the interpretation of the MRI images with Flex-M coil and body coil was not statistically significant, and kappa values signified good concordance between the readers (Table 2).

Table 2

Interobserver variability according to different imaging findings

Variableκ value*P value
Internal opening imaging with Flex-M coil0.4750.002
Internal opening imaging with body coil0.933< 0.001
External opening imaging with Flex-M coil0.635< 0.001
External opening imaging with body coil0.763< 0.001
Secondary tract visualization with Flex-M coil1.000< 0.001
Secondary tract visualization with body coil1.000< 0.001

* Kappa value shows the concordance between Reader 1 and Reader 2

According to Park’s classification, it was seen that 16 (51.6%) patients had intersphincteric fistulas, 14 (45.2%) had transsphincteric fistulas, and one (3.2%) had suprasphincteric fistulas. No extrasphincteric fistulas were detected.

According to St. James University Hospital classification, 13 (41.9%) patients had grade I fistulas, three (9.7%) had grade II fistulas, eight (25.8%) had grade III fistulas, six (19.4%) had grade IV fistulas, and one (3.2%) had a grade V fistula.

Discussion

In this study, we found that MRI could clearly depict perianal fistulas, abscess, secondary tracts, and their connections with sphincters. According to our results, the conspicuity of fistulas was better with fat-suppressed T2-weighted images obtained by MRI with superficial Flex-M coil, compared to those with body coil. This difference was more remarkable in showing the internal opening of the fistula tract (Figures 1 and 2). Besides this, there was excellent inter-observer concordance, which underlined that the operator dependency was low (Table 2).

Figure 1

Twenty-nine-year-old female patient with recurrent fistula. Grade III transsphincteric fistula was seen on axial magnetic resonance images. Axial fat-saturated T2-weighted images obtained with body (A) and Flex-M coils (B). Internal opening was clearer in the image on the right side (arrow)

/f/fulltexts/PJR/38649/PJR-84-38649-g001_min.jpg
Figure 2

Forty-one-year-old male patient with grade I intersphincteric fistula. Axial fat-saturated T2-weighted images obtained with body (A) and Flex-M coils (B). Internal opening was clearly depicted on right picture (arrow)

/f/fulltexts/PJR/38649/PJR-84-38649-g002_min.jpg

Inadequate and/or false evaluation of the internal and external openings, primary and secondary tracts, unexpected anatomy, and concomitant inflammatory bowel disease are among the possible causes of surgical treatment failure [21]. It has been widely expressed that recurrence is inevitable if the correct internal opening is not identified and dealt with.

There are various imaging modalities for the evaluation of perianal fistulas. All these imaging techniques have been compared in various studies. For example, Buchanan et al. compared the accuracy of digital examination, endosonography, and MRI in the preoperative assessment of perianal fistula. In their study, they concluded that endosonography was superior to digital examination, but MRI remained superior in all aspects [22]. Besides this, MRI was shown to be very helpful for physicians in the establishment of a surgery plan before anal fistula operation [18] and in reducing relapses by up to 75% in complex cases [23]. Other techniques such as fistulography and computed tomography (CT) proved no better than clinical examination [20-24]. Fistulography has been shown to be inaccurate sometimes [25], and CT has been shown to be hindered by tissue contrast [26].

Anal endosonography is superior to clinical examination [27], but it has an inadequate field of view (FOV) and poor patient tolerance. Alabiso et al. compared the performances of 3D-endoanal ultrasonography and MRI in the perianal fistulas of patients with Crohn’s disease [28]. In their study, they proposed that MRI was preferable for the detection of supra- and extra-sphincteric fistulas and 3D-USG was preferable in the evaluation of inter-sphincteric fistulas. Significant operator dependence and patient discomfort are known to be important disadvantages of endoanal USG. CT fistulography is another imaging modality, but it is not routinely suggested in anal fistula imaging because it has some limitations in the differentiation of fistula tract, fibrosis, and sphincter muscles [24].

Although each imaging modality has its advantages and disadvantages, MRI is suggested by many experts as the gold standard method for perianal fistula imaging [29,30]. MRI is superior to other imaging techniques due to its excellent soft tissue contrast, operator independence, multiplanar capabilities, and superior FOV [31]. Accuracy rate of MRI in perianal fistula disease differs between studies and ranges between 86 and 100% [24,32-35].

The optimal MRI technique for perianal fistula imaging is still a matter of debate. In a recent study, it was shown that short-tau inversion recovery (STIR) sequences could be a valid alternative to post-contrast T1-weighted sequences in the MRI of perianal fistulas in Crohn’s disease [1]. In another study, Baik et al. compared T2-weighted imaging (T2W), diffusion-weighted imaging (DWI), and contrast-enhanced T1-weighted imaging (CE-T1W), and according to their findings, they concluded that although DWI and CE-T1W improved fistula conspicuity they had comparable diagnostic performance with T2W [5]. Hori et al. investigated the additional value of DW MRI to T2W in the evaluation of perianal fistula disease in comparison with CE-T1W. In their study, they defined DWI as a useful sequence and proposed that it could be a helpful addition to T2W [10]. Due to high spatial resolution, endoanal coils were used in MR imaging as an alternative to body coils for better perianal fistula visualisation. However, surgical concordance remained at 68% for endoanal imaging, whereas it was 96% for MRI with conventional body coil imaging [36]. Patient discomfort and FOV limitation were among the most significant limitations of endoanal imaging.

Pelvic phased-array coils are preferred, because it has been demonstrated in recent studies that they provide higher resolutions. MRI acquisition quality has been improved by multichannel external phased-array coils. With these coils the acquisition time is decreased, signal-to-noise ratio is increased, and larger fields of view are provided without the need for endoanal coils [32]. Because of these advantages, we thought to examine whether using a superficial coil in routine MRI would give better results than conventional MRI with body coil.

Our study does have some limitations. The number of patients was relatively low, and only axial FS T2W images were repeated with Flex-M coil because it would prolong the examination time. Thus, comparison of the coils could only be made based on the T2W images.

As a result, we showed that MRI with Flex-M coil gave clearer images and yielded additional data compared to perianal fistula imaging by MRI with body coil. Relatively easy accessibility and low cost are thought to be the most important advantages of Flex-M coil, making it preferable in perianal fistula imaging.

In conclusion, because proper visualisation of perianal fistulas is very important in the establishment of the treatment plan, we believe that our findings are crucial and applicable to routine daily clinical practice. MRI with Flex-M coil improves imaging quality in perianal fistula imaging and can easily be used in routine daily clinical practice.

Conflict of interest

The authors report no conflict of interest.

References

1 

Lo Re G, Tudisca C, Vernuccio F, et al. , authors. MR imaging of perianal fistulas in Crohn’s disease: sensitivity and specificity of STIR sequences. Radiol Med. 2016. 121:p. 243–251

2 

Seow-Choen, Phillips RK , authors. Insights gained from the management of problematical anal fistulae at St. Mark’s Hospital, 1984-88. Br J Surg. 1991. 78:p. 539–541

3 

Sainio P , author. Fistula-in-ano in a defined population: incidence and epidemiological aspects. Ann Chir Gynaecol. 1984. 73:p. 219–224

4 

Lilius HG , author. Fistula-in-ano, an investigation of human foetal anal ducts and intramuscular glands and a clinical study of 150 patients. Acta Chir Scand Suppl. 1968. 383:p. 7–88

5 

Baik J, Kim SH, Lee Y, et al. , authors. Comparison of T2-weighted imaging, diffusion-weighted imaging and contrast-enhanced T1-weighted MR imaging for evaluating perianal fistulas. Clin Imaging. 2017. 44:p. 16–21

6 

Beckingham IJ, Spencer JA, Ward J, et al. , authors. Prospective evaluation of dynamic contrast enhanced magnetic resonance imaging in the evaluation of fistula in ano. Br J Surg. 1996. 83:p. 1396–1398

7 

Spencer JA, Ward J, Beckingham IJ, et al. , authors. Dynamic contrast-enhanced MR imaging of perianal fistulas. AJR Am J Roentgenol. 1996. 167:p. 735–741

8 

Yıldırım N, Gökalp G, Ozturk E, et al. , authors. Ideal combination of MRI sequences for perianal fistula classification and the evaluation of additional findings for readers with varying levels of experience. Diagn Interv Radiol. 2012. 18:p. 11–19

9 

Ziech ML, Lavini C, Bipat S, et al. , authors. Dynamic contrast-enhanced MRI in determining disease activity in perianal fistulizing Crohn disease: a pilot study. AJR Am J Roentgenol. 2013. 200:p. W170–W177

10 

Hori M, Oto A, Orrin S, et al. , authors. Diffusion-weighted MRI: a new tool for the diagnosis of fistula in ano. J Magn Reson Imaging. 2009. 30:p. 1021–1026

11 

Bakan S, Olgun DC, Kandemirli SG, et al. , authors. Perianal fistula with and without abscess: assessment of fistula activity using diffusion-weighted magnetic resonance imaging. Iran J Radiol. 2015. 12:p. e29084

12 

Cavusoglu M, Duran S, Sözmen CD, et al. , authors. Added value of diffusion-weighted magnetic resonance imaging for the diagnosis of perianal fistula. Diagn Interv Imaging. 2017. 98:p. 401–408

13 

Yoshizako T, Wada A, Takahara T, et al. , authors. Diffusion-weighted MRI for evaluating perianal fistula activity: feasibility study. Eur J Radiol. 2012. 81:p. 2049–2053

14 

deSouza NM, Gilderdale DJ, Coutts GA, et al. , authors. MRI of fistula-in-ano: a comparison of endoanal coil with external phased array coil techniques. J Comput Assist Tomogr. 1998. 22:p. 357–363

15 

Waniczek D, Adamczyk T, Arendt J, et al. , authors. Direct MRI fistulography with hydrogen peroxide in patients with recurrent perianal fistulas: a new proposal of extended diagnostics. Med Sci Monit. 2015. 21:p. 439–445

16 

Ozkavukcu E, Haliloglu N, Erden A , authors. Frequencies of perianal fistula types using two classification systems. Jpn J Radiol. 2011. 29:p. 293–300

17 

Haliloglu N, Gulpinar B, Ozkavukcu E, et al. , authors. Typical MR imaging findings of perianal infections in patients with hematologic malignancies. Eur J Radiol. 2017. 93:p. 284–288

18 

Beets-Tan RG, Beets GL, van der Hoop AG, et al. , authors. Preoperative MR imaging of anal fistulas: does it really help the surgeon? Radiology. 2001. 218:p. 75–84

19 

Parks AG, Gordon PH, Hardcastle JD , authors. A classification of fistula-in-ano. Br J Surg. 1976. 63:p. 1–12

20 

Morris J, Spencer JA, Ambrose NS , authors. MR imaging classification of perianal fistulas and its implications for patient management. Radiographics. 2000. 20:p. 623–635

21 

Abou-Zeid AA , author. Anal fistula: Intraoperative difficulties and unexpected findings. World J Gastroenterol. 2011. 17:p. 3272–3276

22 

Buchanan G, Halligan S, Bartram CI, et al. , authors. Clinical examination, endosonography, and MR imaging in preoperative assessment of fistula in ano: comparison with outcome-based reference standard. Radiology. 2004. 233:p. 674–681

23 

Buchanan G, Halligan S, Williams A, et al. , authors. Effect of MRI on clinical outcome of recurrent fistula-in-ano. Lancet. 2002. 360:p. 1661–1662

24 

Halligan S, Jaap S , authors. Imaging of fistula in ano. Radiology. 2006. 239:p. 18–33

25 

Kuijpers HC, Schulpen T , authors. Fistulography for fistula in ano: is it useful? Dis Colon Rectum. 1985. 28:p. 103–104

26 

Halligan S , author. Imaging fistula in ano. Clin Radiol. 1998. 53:p. 85–95

27 

Sahni VA, Ahmad R, Burling D , authors. Which method is best for imaging of perianal fistula? Abdom Imaging. 2008. 33:p. 26–30

28 

Alabiso ME, Iasiello F, Pellino G, et al. , authors. 3D-EAUS and MRI in the activity of anal fistulas in Crohn’s disease. Gastroenterol Res Pract. 2016. 2016:p. 1895694

29 

Gecse KB, Bemelman W, Kamm MA, et al. , authors. World Gastroenterology Organization, International Organisation for Inflammatory Bowel Diseases IOIBD, European Society of Coloproctology and Robarts Clinical Trials; World Gastroenterology Organization International Organisation for Inflammatory Bowel Diseases IOIBD European Society of Coloproctology and Robarts Clinical Trials. A global consensus on the classification, diagnosis and multidisciplinary treatment of perianal fistulising Crohn’s disease. Gut. 2014. 63:p. 1381–1392

30 

Sahni VA, Ahmad R, Burling D , authors. Which method is best for imaging of perianal fistula? Abdom Imaging. 2008. 33:p. 26–30

31 

Gage KL, Deshmukh S, Majura KJ, et al. , authors. MRI of perianal fistulas: bridging the radiological-surgical divide. Abdom Imaging. 2013. 38:p. 1033–1042

32 

Sheedy SP, Bruining DH, Dozois EJ, et al. , authors. MR Imaging of perianal Crohn disease. Radiology. 2017. 282:p. 628–645

33 

Lunniss PJ, Barker PG, Sultan AH, et al. , authors. Magnetic resonance imaging of fistula-inano. Dis Colon Rectum. 1994. 37:p. 708–718

34 

Maccioni F, Colaiacomo MC, Stasolla A, et al. , authors. Value of MRI performed with phased-array coil in the diagnosis and pre-operative classification of perianal and anal fistulas. Radiol Med (Torino). 2002. 104:p. 58–67

35 

Lunniss PJ, Armstrong P, Barker PG, et al. , authors. Magnetic resonance imaging of anal fistulae. Lancet. 1992. 340:p. 394–396

36 

Halligan S, Bartram CI , authors. MR imaging of fistula in ano: are endoanal coils the gold standard? AJR Am J Roentgenol. 1998. 171:p. 407–412

Copyright: © Polish Medical Society of Radiology This is an Open Access article distributed under the terms of the Creative Commons Attribution-Noncommercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0). License allowing third parties to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially.
 
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.